Dakota CMXDL Dual Element Corrosion Thickness Gauge
SKU:
CMXDL
$0.00
Description

The Dakota Ultrasonics CMXDL is a versatile measurement device designed for efficiently assessing coating and wall thickness. It operates from a single side and offers various modes for different measurement scenarios:
Five user-selected operating modes are included:
Coating On: Measures wall thickness using Pulse-Echo. It separately displays wall & coating thickness.
Coating Off: Measures wall thickness using Pulse-Echo. It displays wall thickness only.
Temperature Compensated: Measures wall thickness using Pulse-Echo and automatically adjusts calibration to compensate for changes in material temperature.
Thru-Coat: Measures wall thickness only using Echo-Echo. (coating is NOT included)
Coating Only Mode (CT): (Coating Thickness) Measures from 0.0005 to 0.100 inches (0.013 to 2.540 mm). Range will vary slightly depending on the coating.
The oversized backlit screen on the CMXDL shows important settings clearly, like Velocity, Operating Mode, Alarms, Scan Mode, etc. You can set the backlight to always on (ON), always off (OFF), or it turns on automatically during measurements (AUTO).
Additional Info
Package Includes
- CMXDL Corrosion Thickness Gauge and Probe
- 4 oz. Bottle of Coupling Fluid
- 2 AA Batteries
- NIST-Traceable Calibration Certificate
- Operating Manual
- RS-232 cable
- Carrying Case
Specifications
| Range in Steel: | Pulse-Echo Mode: Pit and Flaw detection measures from 0.025 - 9.999inches (0.63 to 254 millimeters) Echo-Echo Mode: Thru Paint & coatings measures from 0.100 - 4.000 inches (2.54 to 102 mm). Range will vary depending on the thick |
| Range in Coating Only Mode | 0.0005" - 0.1000" / Resolution: 0.0001" (0.013 - 2.540 mm / Resolution: 0.001 mm) |
| Max Coating | 0.100" |
| Resolution: | .001 inches (0.01mm) |
| Velocity Range: | .0492 to .3936 in./ms 1250 to 9999 meters/sec |
| Units: | Inch or mm |
| Measurement Modes: | - Pulse-Echo (corrosion mode) - Pulse-Echo Coating (Corrosion w/o coating) - Echo-Echo (thru-paint) - Temp Comp (auto adjust for material temp) - Coating Only (Ultrasonic Coating Gauge) |
| Transducer Types: | Dual Element (1 to 10 MHz). |
| Output | RS-232 for upload/download of setups |
| Memory capacity: | TI-CMX: None TI-CMXDL: over 200,000 measurements |
| Power Source: | Three 1.5V alkaline or 1.2V NiCad AA cells |
| Battery Life: | Typically operates for 150 hours on alkaline and 100 hours on NiCad |
| Auto Power off: | After 5 minutes of non-use |
| Display: | 1/8 in. VGA grayscale display 62 x 45.7mm |
| Keyboard: | Membrane switch with twelve tactile keys |
| Case: | Extruded aluminum body with nickel-plated aluminum end caps (gasket sealed). |
| Temp. Limits | Ambient: -20 to 120° F (-30 to 50° C) Material: -0 to 200° F (-20 to 100° C) Special high temperature model TI-CMXDL-H is available for 300°F/150°C and optional probes are offered for higher surface temperatures |
| Weight, net: | 383 grams |
| Dimensions: | 63.5 W x 165 H x 31.5 D mm |
| Warranty: | 2 year limited |
| Certification: | CE Approved, Factory calibration traceable to national standards |
Features
- Built-in memory with RS-232 output for transferring data to a Printer or PC
- Includes Datacomm Data Transfer & Analysis Software and RS-232 Cable, FREE
- Data Files can be user-set for a "grid type" structure where data is formatted in Rows & Columns or a "sequential" file where all data appears in a single column
- Each "grid type" File can be set for up to 999 rows by 52 columns
- Each "sequential" File can be set for up to 512 rows by 1 column
- Each File can be set for "data only" or as "data plus B-scan Graphic". Memory Capacity is over 225,000 data values or over 16,000 data values with B-scan
- Auto probe recognition with Auto Zero or Manual Zero
- Wide variety of dual-element probes can be used for nearly any application.
- Adjustable Gain in 5 steps (VLow, Low, Med, Hi & VHi) in 2db increments providing a 8 db gain adjust range (42–50 db)
- 64 custom setups can be stored and retrieved for convenient use and labeled for easy identification