The new Fluke 50 Series II contact thermometers offer fast response and laboratory accuracy (0.05% + 0.3°C) in a rugged, handheld test tool.
Includes:
Impact absorbing holster
Two 80PK-1 beaded probe thermocouples
Features
Relative time clock on MIN, MAX, and AVG provides a time reference for major events
Electronic Offset function allows compensation of thermocouple errors to maximize overall accuracy
Measures J, K, T, and E-types of thermocouples
Readout in ºC, ºF, or Kelvin (K)
Splash and dust resistant case protected by impact absorbing holster
User-friendly front panel is easy to set up and operate
Sleep mode increases battery life; typical 1000-hour battery life
Battery door allows easy battery replacement without breaking the calibration seal
Optional ToolPak accessory allows the thermometer to hang from any metal object (with the rare earth magnet)or secure around a pipe (with hook-and-loop straps) for hands-free operation.
3 year warranty
Additional thermocouple types R, S, and N (for a total of 7 different types)
Recall function allows logged data to be easily reviewed on the meter display
IR communication port allows data to be exported to optional FlukeView Forms® Temperature PC software for further analysis and graphing
Data Logging up to 500 points of data with user adjustable recording interval
Specifications
| Specifications |
| Temperature accuracy |
| Above -100 °C: |
J, K, T, E, and N-type: ±[0.05% + 0.3°C]* |
|
R and S-type: ±[0.05% + 0.4°C]* |
|
|
| Below -100 °C: |
J, K, E, and N-types: ±[0.20% + 0.3°C]* |
|
T-type: ±[0.50% + 0.3°C] |
|
| |
| Temperature Range |
| J-type: |
-210 °C to 1200 °C |
| K-type: |
-200 °C to 1372 °C |
| T-type: |
-250 °C to 400 °C |
| E-type: |
-150 °C to 1000 °C |
| N-type: |
-200 °C to 1300 °C * |
| R and S-type: |
0 °C to 1767 °C * | |
| Temperature scale |
|
| Applicable standards |
|
| Display Resolution |
|
0.1 °C, 0.1 K < 1000 |
|
1°C, 1 K >=1000 | |
| Note |
| * |
Only the Fluke Models 53 II and 54 II thermometers are capable of measuring N, R, and S-type thermocouples. | |
| Environmental Specifications |
| Operating Temperature |
|
| Storage Temperature |
|
| Humidity (Without Condensation) |
|
0% to 90%; 0 °C to 35 °C |
|
0% to 70%; 0 °C to 50 °C | |
| Safety Specifications |
| Overvoltage category |
|
CSA C22.2 No. 1010.1 1992; EN 61010 Amendments 1,2 | |
| Agency Approvals |
|
| Mechanical & General Specifications |
| Size |
|
173 mm L x 86 mm W x 38 mm | |
| Weight |
|
| Batteries |
|
3 AA batteries; typical 1000-hour life | |